ADF/SCN Unit - After replacement

  1. Turn the power ON.
  2. Press [Execution].
  3. Set the A4-DS chart on ADF tray and press [Execute].

    ADF Scan Position/ Scaling Calibration (600dpi) is performed.
  4. Set the A4-DS chart on ADF tray again, and press [Execute].

    ADF Scan Position/ Scaling Calibration (300dpi Continuous feed) is performed.
  5. Set the A4-DS chart on ADF tray again, and press [Execute].

    ADF Scan Position/ Scaling Calibration (300dpi Single feed) is performed.
  6. Press [Execution].


  7. Press [OK] to complete the menu.