Initialization PW Sensor [B14]
- Start the service support mode. (Service mode startup and operating procedures)
- Select “Adjustment Main Unit: MENU” from the service support mode top menu.
Select “Initialize-PW Sensor”.
Check Point / チェックポイントSelect and execute the corresponding adjustment item using either procedure below.
Selecting by entering the program number
- Select “Input B No.-Main Unit Adjust” from the Adjustment menu.
- Press the [#] button and software number key is displayed on LCD. Enter program number “14”.
- Press the [#] button.
Selecting from menu
- Select “Individual Adjustment: MENU” from the Adjustment menu.
- Select “Initial Setting Menu” from the Individual Adjustment menu.
- Select “Initialize-PW Sensor” from the Initialize menu.
- Initialization is executed.
- The initialization end screen appears. Press the [#] button to return to the menu screen.