Initialization PW Sensor [B14]

  1. Start the service support mode. (Service mode startup and operating procedures)
  2. Select “Adjustment Main Unit: MENU” from the service support mode top menu.
  3. Select “Initialize-PW Sensor”.

    Check Point / チェックポイント

    Select and execute the corresponding adjustment item using either procedure below.

    • Selecting by entering the program number

    1. Select “Input B No.-Main Unit Adjust” from the Adjustment menu.
    2. Press the [#] button and software number key is displayed on LCD. Enter program number “14”.
    3. Press the [#] button.
    • Selecting from menu

    1. Select “Individual Adjustment: MENU” from the Adjustment menu.
    2. Select “Initial Setting Menu” from the Individual Adjustment menu.
    3. Select “Initialize-PW Sensor” from the Initialize menu.
  4. Initialization is executed.
  5. The initialization end screen appears. Press the [#] button to return to the menu screen.