PE DETECTOR ERROR (ERROR CODE: 030001)
Cause
Occurs when PE detection can not be performed correctly due to failure of PE sensor etc.
Suspected cause
- PW sensor failure
- Main Board failure
Troubleshooting
| Step | Confirmation and action | Yes | No |
|---|---|---|---|
| 1 | Check the PE sensor cable connection state. |
Go to step 2 | Reassemble |
| 2 | Check the operation after reinstalling the Main board Replace the main board with normal product and check operation. Does the error occur? |
Replace the PW sensor |
Replace the main board |